Materials required: ISE Detergent Time: 5 minutes Analyzer mode: Manual operation |
|
Wear personal protective equipment. |
ISE Detergent Solution is automatically run through the ISE module as part of the shutdown wash procedure (WASH2). Manually perform additional ISE washes (described in the following procedure) once per shift under either of the following circumstances:
· Dialysis samples are run routinely.
· The system is run more than 8 hours per day.
NOTE
Do not perform the ISE electrode wash more than 3 times per day (once as part of the shutdown wash and twice on a per-shift basis). Pour fresh ISE Detergent into a cup, not a tube, before each wash from the CTT.
At the Menu Panel, select Maint., then select ISE Operation.
At the ISE Operation window, in the Period.wash area, select OFF, then select Set.
In the Wash Electrode area do the following:
Type a CTT position number in the Detergent posi. field, indicating where to place the ISE Detergent container.
In the Container field, select the type of container for the wash solution.
The recommended type of container is 6 : 2mlCUP/Adp.
Pour ISE Detergent Solution into the container and place it in the CTT position entered in step 3a.
In the Wash Electrode area, select Execute, then select Yes when prompted.
At the ISE Operation window, in the Period.wash area, select ON, then select Set.
Close the window, then select Yes when prompted.