ISE Maintenance - Once a day at Shutdown

Washing the electrodes

Materials required

·  ISE Detergent Solution, B01-4174-01

·  10-ml tube

Biohazard

Wear personal protective equipment.
Use universal precautions.

Time: 5 minutes

Analyzer mode: Manual operation

 

NOTE

Executing Wash2 will automatically perform the electrode wash by default. This default can be modified on the ISE parameters of Setting window.

To wash the electrodes (manual wash)

  1. At the Menu Panel, select Maint., then select ISE Operation.

  2. Enter the CTT position number of the ISE detergent container in the Detergent position area. The recommended CTT position number is 15.

  3. In the Container field, select the type of container for the wash solution. The recommended type of container is a 10-ml tube.

  4. Put the ISE detergent solution in the container and place it on the CTT position you entered.

  5. Select Execute to the right of the word Wash Electrode.

  6. Select Exit.