Materials required · ISE Detergent Solution, B01-4174-01 · 10-ml tube |
|
|
Wear personal protective equipment. |
||
Time: 5 minutes Analyzer mode: Manual operation |
|
NOTE
Executing Wash2 will automatically perform the electrode wash by default. This default can be modified on the ISE parameters of Setting window.
At the Menu Panel, select Maint., then select ISE Operation.
Enter the CTT position number of the ISE detergent container in the Detergent position area. The recommended CTT position number is 15.
In the Container field, select the type of container for the wash solution. The recommended type of container is a 10-ml tube.
Put the ISE detergent solution in the container and place it on the CTT position you entered.
Select Execute to the right of the word Wash Electrode.
Select Exit.