Time: 10 minutes Analyzer mode: READY |
Wear personal protective equipment. |
Remove the wash solution reagent containers from RTT1 and RTT2, positions 47 – 50.
Replace the containers with new ones or clean the old containers with DI water.
Refill the containers with fresh solutions as specified in the table below.
RTT1/2 Position |
Wash Solution |
47 |
Probe Wash 1 |
48 |
Probe Wash 2 |
49 |
10% Cuvette Wash |
50 |
DI Water |