Monthly maintenance

Cleaning or replacing the wash solution reagent containers (47 – 50)

Materials required

Biohazard

Time: 10 minutes

Analyzer mode: READY

Wear personal protective equipment.
Use universal precautions.

  1. Remove the wash solution reagent containers from RTT1 and RTT2, positions 47 – 50.

  2. Replace the containers with new ones or clean the old containers with DI water.

  3. Refill the containers with fresh solutions as specified in the table below.

RTT1/2 Position

Wash Solution

47

Probe Wash 1

48

Probe Wash 2

49

10% Cuvette Wash

50

DI Water