Time: 10 minutes Analyzer mode: READY |
Wear personal protective equipment. |
|
Remove the wash solution reagent containers from RTT1 and RTT2, positions 47 – 50.
Replace the containers with new ones or clean the old containers with DI water.
Refill the containers with fresh solutions as specified in the table below.
|
RTT1/2 Position |
Wash Solution |
|
47 |
Probe Wash 1 |
|
48 |
Probe Wash 2 |
|
49 |
10% Cuvette Wash |
|
50 |
DI Water |