Time:5 minutes Analyzer mode: READY |
Wear personal protective equipment. |
Cuvette covers are installed around the probes to prevent water and reagent from entering the dilution and reaction cuvettes.
If there is any splattering on the cuvette covers (1), clean the covers.
Using lint-free towels moistened with deionized water, wipe down the covers.
WARNING
To avoid contamination, do not touch probes or mixing rods.
If splattering is extensive or enters the cuvettes, call your local technical support provider or distributor.