Time: 10 minutes Analyzer mode: Ready |
Wear personal protective equipment. |
|
1 ISE buffer bottle 2 RRV (Reaction) bath oil bottle 3 Isotonic saline diluent bottle 4 Cuvette detergent bottle 5 Cell conditioner bottle |
Caution
Do not attempt to clean the RRV bath oil bottle (2) with water; RRV bath oil and water do not mix.
Unscrew the filter cap at the front of the RRV bath oil bottle, and pull up the tube with the filter.
Disconnect the RRV bath oil bottle level sensor connector. Turn the connector counter-clockwise and pull it out.
Caution
To avoid mixing up the fluid bottles, make a note of the bottle position on the shelf.
Remove the RRV bath oil bottle (2).
Caution
To avoid damaging the level-sensor connector, ensure that it does not get wet.
Refill the bottle with RRV (reaction) bath oil.
Replace the bottle on the shelf in the cabinet.
Connect the RRV bath oil bottle level sensor connector. Push the connector in and turn clockwise.
Insert the filter and hose and fasten the cap.
Note
Make sure that the filter holder is located at the bottom of the bottle.