To assign a blank channel for serum, enter the following information at the Blank Reagent Settings window:
|
No |
Blank CH item |
KV |
BLK item |
STD item |
|||
|
|
No |
Test Name |
|
No |
Test
|
No |
Test Name |
|
1 |
16 |
Fe |
1 |
17 |
Fe-B |
|
|
|
2 |
17 |
Fe-B |
|
|
|
16 |
Fe |
NOTE
The Fe and Fe-B test numbers might be different on your system.
Line 1 Fe-B is assigned as the blank channel for Fe.
Line 2 The Fe calibration sample absorbance is assigned for the calibration of the Fe-B channel.
· To avoid erroneous results do not enter Fe in the STD item column for Line 1.
· To avoid erroneous results do not enter Fe-B in the BLK item column for Line 2.
· Analyze the BLK and STD samples for Fe and Fe-B.
· The sample results are calculated as follows:


NOTE
The KV for Fe-B is not used in the calculations.