Example A: Assigning a blank channel for serum iron

To assign a blank channel for serum, enter the following information at the Blank Reagent Settings window:

No

Blank CH item

KV

BLK item

STD item

 

No

Test Name

 

No

Test
Name

No

Test Name

1

16

Fe

1

17

Fe-B

 

 

2

17

Fe-B

 

 

 

16

Fe

NOTE

The Fe and Fe-B test numbers might be different on your system.

Explanation of the Entries

Line 1 Fe-B is assigned as the blank channel for Fe.

Line 2 The Fe calibration sample absorbance is assigned for the calibration of the Fe-B channel.

Operational Notes

·   To avoid erroneous results do not enter Fe in the STD item column for Line 1.

·   To avoid erroneous results do not enter Fe-B in the BLK item column for Line 2.

·   Analyze the BLK and STD samples for Fe and Fe-B.

·   The sample results are calculated as follows:

image\ebx_1884113806.gif

image\ebx_-2061948287.gif

NOTE

The KV for Fe-B is not used in the calculations.