Calibrating the Direct Bilirubin method using the Total Bilirubin calibration sample and to assign a bilirubin blank, enter the following information at the Blank Reagent Settings window.
No |
Blank CH item |
KV |
BLK item |
STD item |
|||
|
No |
Test Name |
|
No |
Test
|
No |
Test Name |
1 |
23 |
T-Bil |
1 |
25 |
Bil-B |
|
|
2 |
24 |
D-Bil |
1 |
25 |
Bil-B |
23 |
T-Bil |
3 |
25 |
Bil-B |
|
|
|
23 |
T-Bil |
NOTE
The T-Bil and D-Bil test numbers might be different on your system.
Line 1 Bil-B is assigned as the blank channel to T-Bil.
Line 2 Bil-B is assigned as the blank channel to D-Bil, and the T-Bil calibration sample absorbance is assigned for calibration of the D-Bil channel.
Line 3 The T-Bil calibration sample absorbance is assigned for calibration of the Bil-B channel.
Do not run a calibration sample for D-Bil even though its Calc. Mthd field is set to STD at the Analytical Parameters (Chemistry) window.
The sample results are calculated as follows:
NOTE
The KV for D-Bil is not used in the calculations.