Prime and wash settings
These settings are used in performing the electrolyte prime and electrolyte wash operations at the time the PRIME, WASH1, WASH2, and WASH3 operations are performed by the main system. As needed, select the arrow, then select an option. PRIME Asynchronous: ISE Prime runs independently of the system PRIME. Prime start (default): ISE Prime runs concurrently with system PRIME. WASH1 and WASH3 Asynchronous: ISE Wash runs independently of the system Wash. Prime start: Washes run during PRIME operation of main system. Electrode wash start: Washes run during WASH1 and WASH3 operations of main system. Electrode wash start + Prime start: Washes run during the PRIME, WASH1, and WASH3 operations on the main system. Defaults WASH1: Asynchronous WASH2 Prime start: Wash runs during PRIME operation of main system. Electrode wash start: Wash runs during WASH2 operation of main system. Electrode wash start + Prime start (default): Wash runs during during the PRIME and WASH2 operations on the main system. Prime times The number of times to prime (priming begins when you select INITIALIZE, PRIME, or WASH in the Operation Panel). Type 0, 1, 2, or 3; the normal setting is 2. Detergent position Sets the CTT position of detergent used to wash the electrode (this begins when you select WASH). The default position is C-15. |