CAUTION
Do not put items on the Daily QC window in a different order than they display in the System List window to avoid incorrect positioning of results on the Print report. Do not modify the order of the test items on the System Test List window after initial setup. Doing so could result in sample already run being associated with the wrong test name.
In the control list, select the control you want to review.
The daily statistics for the selected control display.
Optional: You can use the daily statistics to establish the Daily QC average and 1 SD values for a control on the Control Data Setup window.
NOTE
For commercial control products, the average and SD values are typically obtained from the package insert for commercial control products. You can use this feature to generate average and SD values for pooled samples used for control purposes.
Select Control Data Registration.
Select the control data.
You can select a specific test or all tests for the current control, a different control or for all controls.
Select OK.
The selected daily control information is sent to the
Control Data Setup window.
To close the QC List dialog box, select Return.