The Active Test List window is only available from the supervisor logon.
Use the Active Test List window to temporarily deactivate a method, so that method-related alarm messages (for example, missing reagents) are not generated. You would use this function for tests that you do not run every day.
After you make your selections, activating or deactivating a method, you must initialize the system for the changes to take effect.
The Active Test List window displays all tests loaded on the system, but the deactivated or unchecked tests are grayed out.
note
At the Active Test List window, you must select all the tests you expect to run. If a test is not selected, the system skips that test. An error message does not display in the Alarm message box on the Operation Panel, but is recorded in the error message log.
· Active is checked if the test is active.
· Test Name displays the name and number of the test.
· R1 Pos No displays the position of R1 on the reagent tray.
· R2 Pos No displays the position of R2 on the reagent tray.