The system stores data for up to 100 calibration/RBL events for each of 2 different reagent lot pairs (R1 and R2), for a total of 200 data sets per method.
Use the Calibration/RBL History window to:
At the Menu Panel, select Calibration, then select Calibration/RBL History.
Select the calibration data to display by selecting Latest or Select:
If the Select option is chosen, then select one of the 3 Test Name options:
Specify a date range, to display data for calibrations performed on specific days for the tests selected:
Type the Start Date (yyyymmdd) of the range (if other than the default, which is the current date).
Type the End Date (yyyymmdd) of the range (if other than the default, which is the current date).
Select Search to display the calibration data for the tests selected for the specified date range.
Note
Select and drag the column heading borders to resize the columns in the calibration data table.
To clear the data for the method selected, select Clear.
When prompted, select Yes.
To reset the calibration select the Reset check box.
NOTE
You cannot cancel the reset once you select Save. If you resize the reset column an save your changes, those changes are retained when you exit out and back into the window.
Select Print, to print the data displayed.
To create a CSV file of the displayed data for archiving, or for exporting the data to another application, do the following:
First, select the data you want to save.
Select Create CSV.
Browse and select a folder in which to save the CSV file.
Select Save.
Close the window.
Use the reset column to override a calibration failure based on the mean of the last 3 passed calibrations.
Select the Refresh checkbox of the most recently failed calibration for a method (lot specific).
The calibration tracking is reset.
Rerun the calibration on the method that failed.
Run Quality Control materials.