On-line Monitor window

Use this window to monitor the exchange of data between the system and the host computer. The monitor displays the data being exchanged and the control signal codes that ensure accurate transmission of information.

You can view up to 300 data items in the window.

Monitoring data exchange between the system and the host

  1. At the Menu Panel, select Maint., then select On-Line Monitor.

The date and time stamp, data, and control signal codes display in the window. Each line represents the sending or receiving of one data item.

If the line begins with S: ADVIA 1800, the system sent the item to the host.

If the line begins with R: ADVIA 1800, the system received the item from the host.

  1. To close the window, select Exit.

Printing on-line monitor information

  1. At the On-Line Monitor window, select Print.

  2. When prompted, select Yes.

Clearing on-line monitor information

  1. At the On-Line Monitor window, select All Clear.

  2. When prompted, select Yes.

Refreshing the on-line monitor information

  1. At the On-Line Monitor window, select Refresh.

The On-Line monitor information is updated.