Use this window to monitor the reagents in the reagent trays (RTT1 and RTT2). The Reagent Inventory window displays the status of all reagents loaded on the reagent trays. The window continuously refreshes with new information, and it updates after each barcode scan.
The Reagent Inventory window automatically opens at start up and remains opened. When using other windows or when selecting the X in the upper right corner, the window minimizes to the taskbar at the bottom.
Note
To set the column back to the default size, select File>Default Column Settings.
The reagents listed in the window are in alphabetical order.
At the Menu Panel, select Reagent, then select Reagent Inventory.
Perform any of the following tasks at this window:
Monitor reagent positions, number of tests, volumes, and days remaining
Print a reagent summary report
Pre-calibrate a new lot of reagents
Close the window.
NOTE
When the Reagent Inventory window opens, the RTT1 reagents display by default. To display the RTT2 reagents, select RTT2.
Notes
·
To run a test even if it is out of range, double-select the test in the QC status fields.
Select the checkbox.
Select OK.
The test displays as OK (green).
NOTE
The system response to an error can be customized.
Check that the analyzer is in READY mode and that the barcode reader is turned on.
Select Barcode Scan.
The Read the Reagent Barcode? message displays.
To execute a barcode scan, select Yes.
The barcode readers on RTT1 and RTT2 scan all positions for each reagent tray and sense the liquid level in each container to determine the volume.
When the Barcode Scan is complete, all information at the Reagent Inventory window refreshes automatically.
NOTE
If you deselect a reagent container while the system is running, the deselect does not take effect until the system returns to the READY state.
To deactivate a reagent container, select the Deselect checkbox next to the applicable Test Name.
When the Deselect box is checked, the system cannot use the reagent. The Test Name is red. The P/B field clears and if more than one of the same reagent is on the reagent trays, the deselected reagent is no longer part of the reagent ranking process.
To reactivate a reagent container, select the Deselect checkbox again.
Select Print RGT Summary.
When the Print Reagent Summary? message box opens, select Yes to print the report.
Three reports print:
·
·
·
Select Total Test Summary.
The Total Test Summary window displays a list of all methods on-board and their total tests counts. The Total Test Summary window only includes paired reagent packs, and packs that have a 'Passed' status for the most recent calibration. If there is only 1 R1 and 1 R2 reagent present, the total tests value is the lower of the tests remaining in the R1 and R2 reagent container. If multiple R1 and R2 reagents are present on the system, the value displayed is the minimum of the sum of the combined R1 reagents and the sum of the combined R2 reagents. For example:
R1a: 20 tests + R1b: 5 tests + R1c: 10 tests = 35 remaining
R2a: 5 tests + R2b: 10 tests + R2c: 10 tests = 25 remaining
Total Tests = 20 tests remaining
To print the Total Test Summary, select Print.
To close the Total Test Summary, select Exit.
Use the Cal Interval Review list to review daily which methods you should calibrate.
note
For barcoded reagents, this information is preset to Siemens guidelines. If this field displays a 0, no Cal Interval Days value shows in the Reagent Inventory window. The operator has an option to modify this filed.
The calibration is reset to its default value only if a full calibration is performed. If an RBL is performed, the Cal Interval Days column is not reset to its default value.
Select Cal Interval Review.
The Cal Interval Review window opens. Instead of an alphabetical list, this window re-sorts the Reagent Inventory list methods in order of Cal Interval Days remaining. Methods with the shortest days remaining are at the top.
To print Cal Interval Review, select Print.
To close the Cal Interval Review, select Exit.
Use this procedure to calibrate the back-up reagent with a new lot number, making it ready for use.
Place the backup (new lot #) reagents on the system along with the primary (old lot #) reagent on the system.
Run a barcode scan.
At the Reagent Inventory window, in the Deselect column, deselect the primary reagent, for example, lot 100.
A check mark displays in the box. This makes the new lot (Lot 200) the primary reagent and Lot 100 the back up.
Run a calibration for the method.
The system only uses the reagents for the calibration that are not deselected (checked).
At the Reagent Inventory window, in the Deselect column, select in the check box next to the old primary reagent. This causes the check mark to be "erased" and the reagent selected again.
Proceed to run the method.
When the system switches from the primary (old) reagent to the back-up reagents, the calibration in use will also switch.
The system follows the rules for multiple reagent containers.