Alarm (result) flags
|
|
|
Displayed on |
|||
Flags |
Flag name |
Reason |
Review/ |
RealTime Monitor |
Patient Report |
Sent to LIS |
* |
Variance |
Photometric tests - Assay data imprecise |
Y |
Y |
N |
|
ISE tests - calibration value error |
N |
Y |
N |
| ||
/ |
Overflow |
Calculation error Do not interpret as a flag. This mark displays in the result field. |
cal error |
cal error |
call error |
|
A |
Clot error |
A clot is detected |
N |
N |
N |
|
B |
Dilution bowl (ISE) |
liquid remained in dilution bowl |
Y |
Y |
N |
|
c |
Calibration mismatch |
Displayed on tests that are in process when a reagent container switch with autocalibration is configured |
Y |
Y |
-- |
|
C |
Calibration error |
Photometric tests - no or faulty calibration |
Y |
Y |
N |
|
ISE tests - ISE calibration range error |
cal error |
cal error |
cal error |
| ||
d |
|
Photometric tests - Abnormal reaction absorbance |
Y |
Y |
N |
|
ISE tests - dilution error, reference electrode value is less than 500 |
N |
Y |
N |
| ||
D |
Absorbance limit (lower) |
Lamp energy low Photometric tests - Abnormal reaction absorbance |
Y |
Y |
N |
|
F |
Temperature error |
The reaction bath temperature is outside the pre-defined range |
Y |
Y |
N |
|
G |
Crash |
A probe crash is detected |
Y |
Y |
N |
|
h, l |
Normal value (upper,lower) |
Y |
Y |
Y |
| |
H, L |
Abnormal value (upper,lower) |
Y |
Y |
Y |
| |
K |
Maximum absorbance limit exceeded |
The maximum absorbance limit is exceeded. NOTE This flag is linked to the Max Limit field at the Analytical Parameters Chemistry window. |
Y |
Y |
|
|
k |
Exceeded the calibration low range |
The calibration low ranges is exceeded. |
Y |
Y |
|
|
j |
Exceeded the calibration highrange |
The calibration low ranges is exceeded. |
Y |
Y |
|
|
M |
Mix error |
A mixer motion error is detected |
Y |
Y |
N |
|
n |
Abnormal number of effective points |
Y |
Y |
N |
| |
N |
Cell blank |
N |
Y |
N |
| |
P |
Prozone |
Failed prozone check |
Y |
Y |
N |
|
Q |
Liquid level sensor error |
A liquid level sensor error occurs |
Y |
Y |
-- |
|
s |
|
Photometric test - insufficient sample |
Y |
Y |
N |
|
|
|
ISE test - insufficient sample |
Y |
Y |
N |
|
S |
Safety |
Photometric test - Displayed after a system error,
for example, lamp energy out of range |
Y |
Y |
N |
|
|
|
ISE test - Displayed after a system error |
Y |
Y |
N |
|
t |
|
Insufficient diluent |
Y |
Y |
N |
|
T |
Thermistor (ISE) |
Abnormal thermistor |
Y |
Y |
N |
|
r |
|
Photometric test - Insufficient reagent |
Y |
Y |
N |
|
R |
Rerun result |
Indicates that a test result is a rerun |
Y |
Y |
Y |
|
u |
|
Photometric tests - Abnormal reaction absorbance |
Y |
Y |
N |
|
U |
Absorbance (upper) |
Lamp energy high Photometric tests - Abnormal reaction absorbance |
Y |
Y |
N |
|