Alarm (result) flags

 

 

 

Displayed on

Flags

Flag name

Reason

Review/
Edit

RealTime Monitor

Patient Report

Sent to LIS

*

Variance

Photometric tests - Assay data imprecise

Y

Y

N

 

ISE tests - calibration value error

N

Y

N

 

/

Overflow

Calculation error

Do not interpret as a flag. This mark displays in the result field.

cal error

cal error

call error

 

A

Clot error

A clot is detected

N

N

N

 

B

Dilution bowl (ISE)

liquid remained in dilution bowl

Y

Y

N

 

c

Calibration mismatch

Displayed on tests that are in process when a reagent container switch with autocalibration is configured

Y

Y

--

 

C

Calibration error

Photometric tests - no or faulty calibration

Y

Y

N

 

ISE tests - ISE calibration range error

cal error

cal error

cal error

 

d

 

Photometric tests - Abnormal reaction absorbance (lower)

Y

Y

N

 

ISE tests - dilution error,

reference electrode value is less than 500

N

Y

N

 

D

Absorbance limit (lower)

Lamp energy low during cell blank measurement check

Photometric tests - Abnormal reaction absorbance (lower)

Y

Y

N

 

F

Temperature error

The reaction bath temperature is outside the pre-defined range

Y

Y

N

 

G

Crash

A probe crash is detected

Y

Y

N

 

h, l

Normal value (upper,lower)

Result exceeds expected range

Y

Y

Y

 

H, L

Abnormal value (upper,lower)

Results exceeds "abnormal" range

Y

Y

Y

 

K

Maximum absorbance limit exceeded

The maximum absorbance limit is exceeded.

NOTE

This flag is linked to the Max Limit field at the Analytical Parameters Chemistry window.

Y

Y

 

 

k

Exceeded the calibration low range

The calibration low ranges is exceeded.

Y

Y

 

 

j

Exceeded the calibration highrange

The calibration low ranges is exceeded.

Y

Y

 

 

M

Mix error

A mixer motion error is detected

Y

Y

N

 

n

Abnormal number of effective points

Insufficient valid data points

Y

Y

N

 

N

Cell blank

Abnormal cell blank

N

Y

N

 

P

Prozone

Failed prozone check

Y

Y

N

 

Q

Liquid level sensor error

A liquid level sensor error occurs

Y

Y

--

 

s

 

Photometric test - insufficient sample

Y

Y

N

 

 

 

ISE test - insufficient sample

Y

Y

N

 

S

Safety

Photometric test - Displayed after a system error, for example, lamp energy out of range or reagent probe crash occurs

Y

Y

N

 

 

 

ISE test - Displayed after a system error

Y

Y

N

 

t

 

Insufficient diluent

Y

Y

N

 

T

Thermistor (ISE)

Abnormal thermistor

Y

Y

N

 

r

 

Photometric test - Insufficient reagent

Y

Y

N

 

R

Rerun result

Indicates that a test result is a rerun

Y

Y

Y

 

u

 

Photometric tests - Abnormal reaction absorbance (upper)

Y

Y

N

 

U

Absorbance (upper)

Lamp energy high during cell blank measurement check

Photometric tests - Abnormal reaction absorbance (upper)

Y

Y

N