Use this window to select the samples for control sample analysis and calibrations, and to temporarily disable a sample so that the rest of a run can proceed.
For example, if you only want to run ISE calibration on serum samples, disable the ISE calibration urine samples. The run proceeds normally, but it does not include urine samples.
At the Menu Panel, select Calibration or QC, then select Sample Select.
Perform any of the following tasks at this window:
Selecting samples available for control analysis calibration
Printing the settings in the Sample Select window
Clearing the settings in the Sample Select window
Close the window.
Select the sample tray to which the sample selections apply.
Select the boxes of the desired samples.
Select samples for other trays, repeat steps 1 - 2.
Select Save.
When prompted, select Yes.
At the Sample Select window, select Print.
At the Print window, select CTT, STT-98, or STT-99 (any combination) to indicate the sample trays you want to print.
Select OK.
The samples on the selected tray(s) print.
At the Sample Select window, select Clear.
At the Clear window, select CTT, STT-98, or STT-99 (any combination) to indicate the range of settings to clear.
Select OK.
All settings for the tray(s) you selected are removed from the window.
note
To override the settings in this window for the next sample run, you can make changes at the temporary Sample Select window that opens after selecting the Start button at the Operation Panel.