At the Calibration Setup window, in the Proc. Test No. list area, locate the applicable method.
In the Blk posi. box for the
method, enter the cup position for the blank solution on the
calibrator/control tray (CTT).
For STD calibrations only, in the STD posi. box, enter the cup position for the calibrator on the calibrator/control tray (CTT).
In the Coeff (FV) box, enter the applicable value.
Note
Do not enter the factor values in the View Calibration Curve window at this time. These values may not be saved. Instead, enter the factor values at the Calibration Setup window.
Select Ctrl/Cal Setup.
Locate the cup positions occupied by the blank solution and the calibrator.
For each item, do the following:
At the Container Type, select the correct tube or sample cup.
In the Meas. Times box, enter the number of aspirations you want taken.
In the Comment box, enter the applicable text that describes the blank or calibrator.
Close the window.
At the window button menu, select Save.
To confirm, select Yes.
Return to Calibration Setup window.