Use the Calibration Setup window to enter the information required to calibrate each photometric test.
You can request automatic recalibration for each test whenever a new reagent container is loaded or after a user-specified time interval. When a time interval expires, the system recalibrates at the start of the next run.
note
You cannot calibrate 2 different reagent lots on the system at the same time. You have to calibrate them one at a time. After the system calibrates the lots separately, you can place both reagents back on the system and run tests.
Log on as supervisor.
At the Menu Panel, select Calibration, then select Calibration Setup.
Perform any of the following tasks at this window:
Select Ctrl/Cal Setup to enter the number of aspirations and container information.
Entering absolute or single-point (STD) calibration methods
Entering multi-standard calibration methods
Printing the calibration setup information.
Deleting the calibration setup information.
Configuring the automatic calibration feature.
Select Save.
To confirm, select Yes.
Close the window.