Calibrating a multi-standard method
In the Proc. test no. list on the left side of the window, locate the applicable method.
Multipoint calibration methods are identified by a Setting
button in the MSTD column
of the Proc.test no. list.
To view the Multi Standard Setup window, select Setting.
In the TT No. area, select 98 or 99.
In the Posi. box of the Blank row, enter the cup position for the blank solution.
For each standard do the following:
Using the information in the calibrator package insert, enter the Lot No., Lot Name, and Exp. Date.
In the Posi. box, enter the cup position of the standard.
This is the location of the calibrator on the tray.
In the Coeff (FV) box, enter the analyte concentration in the standard.
You can get this information from the calibrator package insert or calculate the value using the standard calculator available on the Method Directory. This directory may already be loaded with the Operator's Guide. From the table of contents on your left, select Methods, then select Methods Directory.
Note
Do not enter the factor values in the View Calibration Curve window at this time. These values may not be saved.
Instead, enter the factor values in the Calibration Setup window.
Select Return.
At the Calibration Setup window, select Save.
Return to Calibration Setup window.