Contamination Settings window

Use this window to set the conditions for preventing contamination of reagent probes and reaction tray cuvettes.

Each condition specifies an action to be taken if a specific interferer contaminates a specific receptor. You can set up to 100 conditions each for reagent probes and RRV cuvettes.

To set contamination prevention conditions

  1. Log on as supervisor or tech_manager.

  2. At the Menu Panel, select Setup, then select Contamination Settings.

important

The default setting to enable cuvette and probe contamination prevention is off. Make sure that you turn on these settings by checking the Cell and Probe boxes in the lower left corner in the Contamination Settings window.

  1. Perform any of the following tasks on this window:

Set conditions for avoiding reagent probe contamination

Set conditions for avoiding RRV cuvette contamination

Set conditions for anti-contamination detergent

Print the conditions

Clear the conditions

  1. Close the window.