Use this window to review and edit reagent information, such as on-system stability, calibration frequency, and fill volumes for barcoded reagents and to enter information for non-barcoded reagents.
The Reagent Inventory and Reagent Container Setting windows use this information.
Log on as supervisor or tech_manager.
At the Menu Panel, select Reagent, then select Reagent Information.
The Reagent Information window lists all tests defined for the system.
· Assay Name is the test method name.
· R-Code is the reagent code for each Siemens barcoded reagent container or a 5 digit user defined code for non-barcoded reagents.
· Days Remaining specifies the on-system reagent stability.
· Comment is a user entered remark or reminder.
· Cal Interval is the number of days that a method is calibrated to run.
note
For barcoded reagents, this information is preset to Siemens guidelines. If this field displays a 0, no Cal Interval Days value shows in the Reagent Inventory window. The operator has an option to modify this filed.
· R1 and R2 Fill Volume are the fill volume in mL for reagent containers R1 and R2.
NOTE
For user defined methods, the Assay Name, R-code, Days Remaining and R1-R2 Fill Volume must be entered correctly.
After system installation, all default information for Siemens defined methods displays in this window.
Update information in this window as follows:
Select in the field you want to change and enter the new information.
At the Save Reagent Information window, select Save.
Select OK.
Close the window.