Use this window to verify (and if needed, to change) container types and measurement frequencies for CTT positions used in the method (they are used to test calibrators, controls, and special diluents).
The settings are normally entered in the Ctrl/Cal Sample Setup window. Those settings apply to all methods specified for that position (the methods for a given CTT position are specified in the Calibration Setup and QC Sample Definition windows).
At the Analytical Parameters (Chemistry) window or the ISE Parameter Settings window, select CTT set.
At the CTT Setting window, verify or change the Cup/Tube Type and the number of measurements (Reps) as needed.
notes
· The tests to perform on each position are displayed in the Test name field.
· Select the Next and Prev buttons to find more tray positions. The Test names box lists the methods assigned to each CTT position.
To save the settings and return to the Analytical Parameters (Chemistry) or ISE Parameter Settings window, select Return.
If you did not make any changes, select Cancel.