Use the Ratio Parameters window to create up to 20 ratios.
The ratio can contain 8 tests.
At the Menu Panel, select Setup, then select Ratio Parameters.
Enter the following information in the Analysis condition area:
In the Analy.cond.no. box, enter the test number (101 to 120) you want to identify the ratio.
In the Sample Type list, select the applicable sample type.
In the Test name box, enter the ratio name (up to 6 alphanumeric characters).
In the Digits box, enter the number of decimal places (1 to 4) you want in the ratio result.
If you selected Do, select Qualit.set.
Select Rerun cond.
Enter the following information in the Standard value set area:
In the Abnormal v. (serum) H box, enter the result value above the triggered H flag (panic high).
In the Abnormal v. (serum) L box, enter the result value below the triggered LH flag (panic low).
To assign the expected range values, select Normal val.set Normal val.set.
Enter the following information in the Real-time Calculation Ratio Parameters area:
In the test x and test y boxes, enter the test numbers for the tests you want to use in the ratio.
The test numbers are provided in the Process.Test list.
In the Factor a through Factor f boxes, enter any numerical factors that are required in the ratio.
In the Formula box, enter the ratio expression.
To test the calculation of the ratio result, select Simulation.
Select Save.
To confirm, select Yes.
Review the process and print assignments for the ratio in the Process Sequence window.