Test Select window

You select tests in this window for patient and control sample analysis and calibrations.

Use this window for the following tasks:

·   Disable ("down") a test.

For example, a test may be downloaded from a host system to analyzers in your lab, but that test is not run on the ADVIA Chemistry System.

·   To "down" a test, select one of the Test Select radio buttons, and deselect it from the corresponding test table.

note

You can remove tests for calibrators and controls from the current run using the Calibration and Control windows, but they stay selected for subsequent runs.

Selecting tests for an analysis

Effect of tests added

  1. At the Menu Panel, select Request, Calibration, or QC, then select Test Select.

  2. At the Test Select window, make your selections:

  1. In the Test Select area, select an assay name:

 

Routine/Interrp.

Ordinary control Sample meas.

Auto control Sample meas.

Auto  control meas. after calib.

Ordinary calibration meas. (BLK)

Ordinary calibration meas. (STD)

Auto calibration meas. (BLK)

Auto calibration meas. (STD)

Special calibration meas. (BLK)

Special calibration meas. (STD)

 

  1. In the Test Table area, select the tests you want to run.

You can set the order in which the tests display in the other windows in the Process Sequence window.

The tests you select display in the Test List on the right.

The Test Table displays 80 tests at a time. To display more tests, select 1-80, 81-160, 161-240, 241-320, or 321-326.

  1. To clear all selections, select Clear Test List.

  2. When prompted, select Yes.

  3. Repeat steps a and b to select tests for each selection in the test select area.

  4. Select Save.

  5. To save your selections, select Yes.

  1. Printing selections:

  1. Select Print.

  2. Print selections.

·   To print selections for all items, select All.

·   To print selections for specified items, select Test select, then select each item you want to print.

  1. Select OK.

  1. Close the window.